• New
Medtronic NIM-Response 2.0 Monitor
zoom_out_map
  • New

Medtronic NIM-Response 2.0 Monitor

05012207
Second Hand
€200.00
Last items in stock

Medtronic NIM-Response 2.0 Monitor

Neural Monitoring System

The Medtronic NIM-Response 2.0 is a four-channel EMG monitor for intraoperative use in surgical procedures where a motor nerve is threatened by inadvertent manipulation. The NIM-Response 2.0 records the electromyographic (EMG) activity of the muscles innervated by the affected nerve.

Real time: 5 Visitor right now
Hurry up! Only 1 item(s) left in Stock!
 

Worldwide Delivery

 

After Sales Service Guaranteed

 

Secured Payment

Medtronic NIM-Response 2.0 Monitor

Neural Monitoring System

The Medtronic NIM-Response 2.0 is a four-channel EMG monitor for intraoperative use in surgical procedures where a motor nerve is threatened by inadvertent manipulation. The NIM-Response 2.0 records the electromyographic (EMG) activity of the muscles innervated by the affected nerve.

This monitor will aid in early nerve identification by providing the surgeon with a tool to help locate and identify the particular nerve at risk in a surgical field. It continuously monitors the EMG activity of the muscles innervated by the nerve at risk to minimize trauma by alerting the surgeon when a particular nerve has been activated.

  • 4-channel EMG monitor that can be selected individually and simultaneously.
  • High contrast digital color touch screen.
  • Increased sensitivity and reduced interference.
  • Printable log of patient's EMG activity.
  • More sensitive to low level EMG activity with better signal to noise ratio.
  • Quick installation.
05012207

Specific References

Second Hand

Dimensions:

- Height: 10.6” (27 cm)

- Width: 12.2” (31 cm)

- Depth: 14.5” (37 cm)

- Weight: 18.7 lbs (8.5 kg)

Display:

- Type: High contrast, digital, graphic Color, visible in complete darkness.

- Resolution: Display 600H x 800W pixels, Touch Panel 256H x 256W

- Dedicated Function Event Touch Screen Controls: For Amplitude, Time Display and Capture.

- Vertical Display: 50, 200, 500 and 2,000 μV display modes.

- Event Capture: Enable/disable capture mode indicator on touch screen.

- Time Scale: 50ms or 10s display modes.

Amplifier

- Channels – 1 to 4: Individually and simultaneously selectable.

- Input Sensitivities: 5 – 10,000 μV peak-to-peak AC-coupled

- Sensitivity Selection: Automatically zeroed

- Bandpass: 100 Hz - 2 kHz EMG Display; 200 Hz - 3.2 kHz Audio EMG Speaker

- Input Noise: 5-11 μV p-p < 3 μV RMS @ DC - 4 Hz, inputs shorted

- Input Impedance: > 200 K Ohm ± 10%

- DC Offset Rejection: ± 0.540 V DC Rejection

- Common Mode Rejection: >90 dB @ 60 Hz, balanced inputs, >70 dB @ 60 Hz, 1 K

- Channel Enable/Disable Controls: Dedicated function touchpads for independent channel enable/disable.

- Event Threshold Control and Display: Adjustable Graduated Touch Bar with Voltage threshold displayed.

- Patient Isolation: 1,000 Vrms 60 Hz < 100 μA

Impedance Measurement

- Control: Automatic CHECK ELECTRODE feature

- Measuring Signal: 20 μA peak-to-peak, 60 Hz Square wave

- Measurement Range: 0-200 K Ohm ± 10%

Artifact Detection and Rejection

- Stimulus Artifact: Synchronized and adjustable muting.

- Electrocautery Interference: Automatic detection, muting and warning.

- Electrode Lead Off: Automatic detection, muting and warning.

Stimulator

- Stimulus Type Constant: Constant Current

- Stimulus Range: 0-30 mA, maximum 120 V compliance

- Stimulus Control: Digitally controlled, range – dependent adjustment increments of .01, .05, 1.0 and 5.0 mA.

- Stimulus Output Accuracy: ± .02 mA (or ± 5% of reading at 1K load) over Stimulus Range.

- Stimulus Adjustment: Graduated Touch Screen Control with display of command current and delivered current.

- Stimulus Measurement Accuracy: ± .02 mA (or ± 5% of reading at 1K load) over Stimulus Range.

- Internal Fuse: 32 mA Type F, 250V 5 x 20mm (It must be Xomed # 8250615, other similar fuses may not give the same degree of protection.

Stimulus Characteristics

- Waveform: Monophasic, square pulse

- Duration: Software selectable, 50, 100, 150, 200 or 250 μs

- Rise Time: Less than 20 μs

- Rate: Software selectable, 1, 4, 7, or 10 Hz

- Stimulus Probe: Monopolar (standard) or bipolar

I / O - Printer Output / Disk Drive Output

- Printer Interface: Hewlett-Packard: Model: DeskJet 5650 printer.

- Connection: Centronics Parallel 36-pin female or USB (2)

Video Output

- Interface: VGA Compatible, 800 x 600 resolution

- Connection: 15-pin HD

Electrical

- Input Voltage: 100V, 120V, 220V, and 230-240V ± 10%

- Frequency: 50/60 Hz

- Power Consumption: 47 W Nominal, <70 W Peak

- Auxiliary AC output: 250 VA Max.