Fresenius Vial Pilote A2 IEC Syringe Pump
Used Fresenius Vial Pilote A2 IEC Syringe Pump
The Pilot A2 is a syringe pump that allows the infusion of intravenous agents at a low and precise flow rate.
The wide choice of syringes, the use of universally known symbols for the controls, the synoptic display of alarms make the Pilote A2 a very simple device to use.
Worldwide Delivery
After Sales Service Guaranteed
Secured Payment
Used Fresenius Vial Pilote A2 IEC Syringe Pump
The Pilot A2 is a syringe pump that allows the infusion of intravenous agents at a low and precise flow rate.
The wide choice of syringes, the use of universally known symbols for the controls, the synoptic display of alarms make the Pilote A2 a very simple device to use.
Adjustable occlusion detection, correct positioning detection and integral syringe protection guarantee an optimum level of safety.
Its technical characteristics, the flow rate range from 0.1 to 400 ml/h (configured at 200 ml/h) and its excellent accuracy (±1% on the device) make the Pilote A2 the ideal device for medical departments.
Characteristics:
- Fast and intuitive programming
- Immediate handling
- Easy readability through the use of pictograms
- Wide choice of sizes and brands of syringes
- Wide range of configurable flow rates
- Memorization of drug names
- Storage of infusion-related parameters
- Long battery life, which can be consulted in hours and minutes on the keyboard
- Infusion pressure management (3 preset levels)
Technical Data Sheet :
Mains and battery operation.
Flow rate ranges: 0.1 to 400 ml/h (200 ml/h default), 0.1 ml/h increment, adjustable flow rate range.
Choice of syringe: 12 types of syringes 50/60 ml and 20 ml, Luer Lock; 10 ml optional, automatic recognition of syringe volume.
Accuracy on the delivered flow rates: accuracy of the device ± 1% /, accuracy on the syringes ± 2%.
Bolus flow rate: selectable from 50 to 500 ml/h with display of the bolus volume infused,
Purge flow rate: 500 ml/h.
Alarms and Security: direct reading, simultaneous processing of several messages:
- Integral protection of the syringe: detection of syringe wings, syringe holding, piston head and antisiphon system, mechanical disengagement.
- infusion control: end of infusion alarm and pre-alarm, occlusion alarm; non validation alarm, end of pause alarm, alarm and limit volume pre-alarm.
- device control: mains disconnection alarm, battery discharge alarm and pre-alarm, battery life display, preventive maintenance; internal malfunction (self-test, rotation, motor), error messages.
Occlusion alarm threshold: 3 preset levels: 300 / 525 / 900 mmHg.
Power supply: 230 V ~ 50/60 Hz, 6 V 1.1/1.3 Ah gel lead battery; autonomy: 7 h minimum at 5 ml/h, external power supply 12-15 V dc 15 W.
Dimensions / Weight: H=12 x W=33 x D=15,5 cm / 2,2 kg.
Specific References